What is the use of ND-DQ9 five-element analyzer?
Product model: ND-DQ9
Product features: At present, most of the five elements in steel can be detected in China are split testing, and the product types of integrated testing are less. Taking the currently popular ND-DQ9 five elements analyzer as an example, the instrument can freely enter or delete the standard sample point when establishing the standard sample curve, repeatedly perform regression processing, and determine the curve choice according to the correlation coefficient. The construction of standard curve completely realizes digitalization and formulation, and gets rid of the traditional backward method of manual curve construction. It can easily call and modify the curve, and carry out man-machine dialogue, so that the scope of use of the instrument has been greatly expanded.
Details of ND-DQ9 Five Elements analyzer:
ND-DQ9 type five elements analyzer uses
The five elements analyzer is used to test the content of manganese, phosphorus, silicon, carbon and sulfur in steel. Manganese, phosphorus, silicon, carbon and sulfur in steel are the most important and basic elements, so we are used to calling them the five elements of steel. According to the current standard GB/T5613-1995 cast steel grade expression method and GB/T5612-1985 cast iron grade expression method, it is explained that the five elements are to distinguish the grade and quality of ordinary steel, and its content directly affects the mechanical properties of steel. Steel and casting enterprises regard the inspection of the five elements of the product as one of the most important inspection.
Development basis
Manganese (according to GB/T 223.63-1988 standard) was oxidized by silver salt - ammonium persulfate spectrophotometry.
Phosphorus element (according to GB/T 223.61-1988 standard) using sodium fluoride - stannous chloride molybdenum blue degree method
Silicon element (according to GB/T 223.5-1997 standard) using ferrous reduction - silicon molybdenum blue degree method
Carbon (according to GB/T 223.69-1997 standard) adopts the gas capacity method
Sulfur element (according to GB/T 223.68-1997 standard) adopts iodine measurement
ND-DQ9 five elements analyzer Technical parameters:
1, gas capacity method liquid absorption fixed carbon, iodine measurement of sulfur, photometric legal manganese, phosphorus, silicon
3, the program built-in steel, iron and other material classification gear, can be selected according to need.
4, imported precision pressure sensor detection data, measurement results digital display direct reading, automatic printing.
5, the program fully automatic operation, data calculation by microcomputer control
6, analysis range: manganese: 0.010-20.000% Phosphorus: 0.0005-1.0000% Silicon: 0.010-20.000%
Carbon: 0.015-12.000% Sulfur: 0.0020-2.0000%
7. Analysis time: about 250s (excluding weighing sample sampling time)
8, analysis error: in line with the national standards GB223.69-8, GB223.68-89 and GB223.3 ~ 5-88 standards
9, power supply voltage: 220V±10% 50Hz
10, power consumption: about 50VA
11, power gas: oxygen, pressure 20 ~ 40KPa
12, ambient temperature: 0℃--40℃, relative humidity: 30%--80%
13. Net weight of the instrument: about 65kg